Dose and image quality in mammography with an automatic beam quality system.

作者信息

Young K C, Ramsdale M L, Rust A

机构信息

Department of Medical Physics, Royal Surrey County & St Luke's Hospitals, Guildford, UK.

出版信息

Br J Radiol. 1996 Jun;69(822):555-62. doi: 10.1259/0007-1285-69-822-555.

Abstract

Radiation dose, contrast and image quality for automatic beam quality selection (OPDOSE) with the Siemens Mammomat 3000 has been investigated for different breast thicknesses and compared with those found using manually set tube potentials and a molybdenum target and filter. Automatic beam quality selection was found to have a negligible effect for breasts with a compressed breast thickness of less than 45 mm. However, for larger breasts substantial dose savings were achieved for a loss in contrast. For mammograms of compressed breast with a thickness in excess of 60 mm the mean glandular dose (MGD) per film was 2.90 mGy for manually selected tube potentials with a molybdenum/molybdenum target filter combination as compared with 1.87 mGy using 26 kVp and a tungsten target with rhodium filtration. The contrast loss in using OPDOSE was measured with a test object to be about 10% for breast thicknesses in excess of 45 mm. The standard breast model, which assumes a 50% glandular content, did not provide a good fit to the MGD for women attending for breast screening in the age range 50 to 64 years.