Schulz-Wendtland R, Hermann K-P, Wenkel E, Böhner C, Lell M, Dassel M-S, Bautz W A
Radiologisches Institut--Gynäkologische Radiologie, Friedrich-Alexander-Universität Erlangen-Nürnberg.
Rofo. 2007 May;179(5):487-91. doi: 10.1055/s-2007-962983.
Reduction of radiation exposure at an adequate image quality by optimizing the radiation quality for a new system of full-field digital mammography using a digital detector (a-Se).
The investigations were performed using a digital mammography system Novation (Siemens, Erlangen). The system was constructed with a bimetal anode (molybdenum and tungsten) and the possibility of changing the filter (molybdenum/rhodium). The test object was the Wisconsin Mammography Random Phantom Model 152 A (Radiation Measurements Inc.) of which images were acquired using the digital technique with the tungsten anode and rhodium filter at different tube voltages (26-35 kV) and tube loads (40-100 mAs) and compared to images in the molybdan/molybdan molybdenum/molybdenum technique. To quantify the image quality, we used the detection rate of the simulated lesions in the phantom.
Increasing the tube voltage significantly decreases the average glandular dose when using AEC (Automatic Exposure Control), i. e., constant detector dose. At the same time, the image quality decreases significantly with respect to the detection rate (26 kV, 1 mGy, 95.1 %; 35 kV, 0.7 mGy, 82.7 %). As a good compromise between the necessary diagnostic image quality and the lowest dose exposition, 28 kV and 60 mAs were selected for imaging with the tungsten/rhodium anode/filter combination. A further change to the tube load did not make sense because a decrease of 10 % resulted in a significant decrease in the detection rate while only a 2 % increase in detection rate was achieved for a 65 % increase in radiation exposure.
The results of this phantom study demonstrate that the routine use of the tungsten anode in combination with a rhodium filter for full-field digital mammography with an a-Se detector in contrast to a molybdan/molybdan molybdenum/molybdenum anode/filter combination results in a reduction of the average glandular dose of up to 30 % without loss of diagnostic image quality.
通过优化使用数字探测器(非晶硅)的新型全场数字化乳腺摄影系统的辐射质量,在保证足够图像质量的同时降低辐射剂量。
使用Novation数字化乳腺摄影系统(西门子,埃尔朗根)进行研究。该系统采用双金属阳极(钼和钨),并可更换滤过器(钼/铑)。测试对象为威斯康星乳腺摄影随机体模模型152A(辐射测量公司),使用数字技术在不同管电压(26 - 35 kV)和管负载(40 - 100 mAs)下,采用钨阳极和铑滤过器采集图像,并与钼/钼技术下的图像进行比较。为了量化图像质量,我们使用了体模中模拟病变的检测率。
使用自动曝光控制(AEC)时,即探测器剂量恒定,增加管电压可显著降低平均腺体剂量。同时,图像质量在检测率方面显著下降(26 kV,1 mGy,95.1%;35 kV,0.7 mGy,82.7%)。作为必要的诊断图像质量和最低剂量暴露之间的良好折衷,选择28 kV和60 mAs用于钨/铑阳极/滤过器组合成像。进一步改变管负载没有意义,因为减少10%会导致检测率显著下降,而辐射暴露增加65%时检测率仅提高2%。
该体模研究结果表明,与钼/钼阳极/滤过器组合相比,在全场数字化乳腺摄影中常规使用钨阳极与铑滤过器结合非晶硅探测器,可使平均腺体剂量降低高达30%,且不损失诊断图像质量。