Wang N, Zhou W, Ye T
Zhongshan Ophthalmic Center, Sun Yat-sen University of Medical Sciences, Guangzhou.
Zhonghua Yan Ke Za Zhi. 1995 Nov;31(6):412-6.
A new computerized method of measurement and analysis of the anterior chamber configuration in the human eye is described. The image or photograph of the anterior chamber taken with a slitlamp microscope is inputted into a computer with digitized camera or digitizer. After cornea, iris and lens surfaces are digitized, methematical corrections are made for the distortion from the slitlamp microscope and the refractive effect of cornea and aqueous humor to convert an optic distorted image into a true image. From the corrected image of anterior chamber, the data and parameter of anterior chamber configuration are measured and calculated. The reproducibility and accuracy of the method were 1.78% approximately 4.66% and 0.05 mm respectively and found to be adequate for clinical study of human eye under physiologic or disease state.
描述了一种用于测量和分析人眼前房形态的新型计算机化方法。用裂隙灯显微镜拍摄的前房图像或照片通过数字化相机或数字化仪输入计算机。在对角膜、虹膜和晶状体表面进行数字化处理后,对裂隙灯显微镜产生的畸变以及角膜和房水的屈光效应进行数学校正,将光学畸变图像转换为真实图像。从前房的校正图像中测量并计算前房形态的数据和参数。该方法的可重复性和准确性分别约为1.78%和4.66%,以及0.05毫米,发现足以用于人眼在生理或疾病状态下的临床研究。