Zhou W
Zhonghua Yan Ke Za Zhi. 1989 Mar;25(2):112-4.
A simple and accurate formula for the measurement of the anterior chamber depth with keratometer and slitlamp is described. The calculation takes into consideration the curvature of the anterior corneal surface and the refractive index of the aqueous. The method was tested with a model of anterior chamber double-blindly and proved accurate and reliable. The average anterior chamber depth of 50 patients (100 eyes) measured by this method was 2.97 +/- 0.22 mm.
本文描述了一种使用角膜曲率计和裂隙灯测量前房深度的简单准确公式。该计算考虑了角膜前表面的曲率和房水的折射率。该方法在前房模型上进行了双盲测试,结果证明准确可靠。用该方法测量的50例患者(100只眼)的平均前房深度为2.97±0.22mm。