AVE微型冠状动脉支架移位。

Migration of the AVE Micro coronary stent.

作者信息

Wong P, Leung W H, Wong C M

机构信息

Cardiac Catheterization Laboratory, Adventist Hospital, Hong Kong.

出版信息

Cathet Cardiovasc Diagn. 1996 Jul;38(3):267-73. doi: 10.1002/(SICI)1097-0304(199607)38:3<267::AID-CCD11>3.0.CO;2-E.

Abstract

The AVE Micro coronary stent is a balloon-expandable stent with a design that is different from the commonly used slotted tubular or coil stents. The stent delivery system is low in profile and very trackable so that it can negotiate tortuous vessels to reach distal lesions. It also can pass through proximally deployed stents easily. Its moderate radioopacity allows precise stent placement. However, as illustrated in the three case reports presented here, the stent struts did not seem to be firmly embedded into the arterial wall after initial deployment, so that stent migration occurred during subsequent passage of a balloon into the stent for poststenting high pressure balloon dilatation. This new phenomenon of stent migration has not been reported previously with other stents.

摘要

AVE微型冠状动脉支架是一种球囊扩张型支架,其设计与常用的开槽管状或线圈状支架不同。支架输送系统外形小巧,追踪性极佳,能够通过迂曲血管到达远端病变部位。它还能轻松穿过近端已植入的支架。其适度的不透X线性便于精确放置支架。然而,正如这里展示的三个病例报告所示,在最初植入后,支架支柱似乎并未牢固地嵌入动脉壁,因此在随后通过球囊进入支架进行支架植入后高压球囊扩张时发生了支架移位。这种新的支架移位现象此前在其他支架中尚未有过报道。