Suppr超能文献

一种新型临时性静脉滤器——Spring滤器的体外评估

In vitro evaluation of a new temporary venous filter: the Spring filter.

作者信息

Xian Z Y, Roy S, Hosaka J, Kuroki K, Kvernebo K, Enge I, Laerum F

机构信息

Institute for Surgical Research, National Hospital (University of Oslo), Norway.

出版信息

Cardiovasc Intervent Radiol. 1995 Sep-Oct;18(5):315-20. doi: 10.1007/BF00203683.

Abstract

PURPOSE

To evaluate in vitro the function efficacy of a new variable-sized, temporary venous filter, the Spring filter (SF).

METHODS

The SF was tested in a flow phantom, using flexible, thin-walled polyethylene tubes 12, 14, and 16 mm in diameter to simulate veins. Clots of three sizes were used: 6 x 10 mm, 6 x 20 mm, 9 x 20 mm. Filter deployment was performed, ensuring that its functional diameter after placement was a predetermined multiple of the radius 'r' of the tube: 3.0r, 2.8r, 2.6r, 2.4r. The terminal coil was visually examined for changes in configuration. Clot retention by the filter for a period of 3 min was recorded as a success. The baseline intraluminal pressure and the maximum pressure reached after clot introduction were recorded. The trapping ability was calculated from the results for 10 clots. The effect of the following parameters on filter function were studied: size of embolus, caliber of simulated vein, and functional diameter of filter.

RESULTS

The filter configuration was stable. Not a single instance of axial tilting occurred on deployment. Functional diameter was the major determinant of filter function. One hundred percent of clots were trapped irrespective of clot size and tube caliber when the functional diameter was 3.0r. Only 73% (95% confidence interval (CI): 57% - 88%) and 77% (95% CI: 61% - 92%) of small emboli were held up by the filter when the functional diameters were 2.6r and 2.4r, respectively (p = 0.0001). A significant reduction in clot trapping was encountered even with medium clots when the functional diameter was 2.4r (p = 0.02).

CONCLUSION

Against the background of available data on retrievable vena caval filters, the current model of SF warrants further investigation.

摘要

目的

在体外评估一种新型可变尺寸临时静脉滤器——弹簧滤器(SF)的功能效果。

方法

在流动模型中对SF进行测试,使用直径为12、14和16毫米的柔性薄壁聚乙烯管模拟静脉。使用三种尺寸的血凝块:6×10毫米、6×20毫米、9×20毫米。进行滤器置入操作,确保置入后其功能直径为管子半径“r”的预定倍数:3.0r、2.8r、2.6r、2.4r。目视检查终端线圈的形态变化。记录滤器在3分钟内对血凝块的截留情况,记为成功。记录基线腔内压力以及引入血凝块后达到的最大压力。根据10个血凝块的结果计算捕获能力。研究以下参数对滤器功能的影响:栓子大小、模拟静脉管径以及滤器功能直径。

结果

滤器形态稳定。置入时未出现一例轴向倾斜。功能直径是滤器功能的主要决定因素。当功能直径为3.0r时,无论血凝块大小和管子管径如何,100%的血凝块被捕获。当功能直径为2.6r和2.4r时,分别只有73%(95%置信区间(CI):57% - 88%)和77%(95%CI:61% - 92%)的小栓子被滤器截留(p = 0.0001)。当功能直径为2.4r时,即使对于中等大小的血凝块,血凝块捕获率也显著降低(p = 0.02)。

结论

在可回收腔静脉滤器现有数据的背景下,当前的SF模型值得进一步研究。