Lorch H, Zwaan M, Kulke C, Weiss H D
Department of Radiology, Medical University of Lübeck, Lübeck, Germany.
Cardiovasc Intervent Radiol. 1998 Mar-Apr;21(2):146-50. doi: 10.1007/s002709900232.
To evaluate the clot trapping capacity of different temporary vena cava filters in a vena cava model.
A vena cava flow model was built using PVC tubing, a hemodialysis membrane and a pulsatile pump. Blood was imitated by a Dextran 40 solution. Five different temporary vena cava filters and two prototypes were tested using human thrombi. The mechanism of clot capture was observed.
Decreasing rank order according to decreasing percentage of clots captured for the 21-mm diameter vena cava model was Cook (C) > Angiocor (A) > Cordis (CD) > Antheor (TF-6) > DIL for thrombi with a diameter of 3 mm and A > C > CD > TF-6 > DIL for 5-mm thrombi. In a cava with diameter of 28 mm, decreasing rank order was C > CD = A > TF-6 > DIL and C > CD = A > DIL > TF-6 for 3- and 5-mm thrombi, respectively. Two new prototypes, the TF-8 and TF-10 filters, achieved better results than the TF-6 filter and were in most conditions comparable to the A and CD filters. In most cases, thrombi were trapped between filter and cava wall.
The vena cava flow model demonstrates significant differences in rates of clot capture (range 22%-98%) depending on cava diameter, thrombus size, and filter type.
在腔静脉模型中评估不同临时性腔静脉滤器的血栓捕获能力。
使用聚氯乙烯管、血液透析膜和脉动泵构建腔静脉血流模型。用右旋糖酐40溶液模拟血液。使用人体血栓对五种不同的临时性腔静脉滤器和两种原型滤器进行测试。观察血栓捕获机制。
对于直径21毫米的腔静脉模型,按捕获血栓百分比递减的顺序排列为:对于直径3毫米的血栓,库克(C)>血管心(A)>科迪斯(CD)>安特奥尔(TF - 6)>迪尔(DIL);对于5毫米的血栓,A>C>CD>TF - 6>DIL。在直径28毫米的腔静脉中,对于3毫米和5毫米的血栓,按捕获血栓百分比递减的顺序分别为:C>CD = A>TF - 6>DIL和C>CD = A>DIL>TF - 6。两种新型原型滤器,即TF - 8和TF - 10滤器,比TF - 6滤器取得了更好的结果,并且在大多数情况下与A和CD滤器相当。在大多数情况下,血栓被困在滤器和腔静脉壁之间。
腔静脉血流模型显示,根据腔静脉直径、血栓大小和滤器类型,血栓捕获率存在显著差异(范围为22% - 98%)。