Lee T K, Egbert T P, Westenskow D R
Department of Anesthesiology, University of Utah, Salt Lake City 84132, USA.
J Clin Monit. 1996 Jul;12(4):293-7. doi: 10.1007/BF02221749.
Noninvasive blood pressure measured from the superficial temporal artery has been shown to correlate well with pressure in the brachial artery. The supraorbital artery may be an even better site for monitoring blood pressure on the forehead because it originates from the internal carotid artery, and it is easier to locate anatomically. This study compares mean pressure measured oscillometrically over the supraorbital artery and at the upper arm.
Oscillometric signals from the supraorbital artery were recorded in 20 surgical patients under general anesthesia using a 2.5- x 1-cm bladder attached to the forehead with a self-adhesive pad. Blood pressure was measured simultaneously from the arm using a Dinamap 1846 blood pressure monitor, and the resulting data compared with the supraorbital artery measurements.
The mean difference between 219 pairs of blood pressure measurements, from the forehead and the arm, was 3.8 mm Hg. The standard deviation of the differences was 7.4 mm Hg. The linear regression equation for the data was y = 0.98x + 3.25, with a standard error of estimate of 7.31 mm Hg. The correlation coefficient between the two measurements was 0.82.
The results show that mean blood pressures measured oscillometrically from the supraorbital and brachial arteries agree and correlate well with each other. The supraorbital artery should be a good alternative site for blood pressure measurement.
已证明从颞浅动脉测量的无创血压与肱动脉血压具有良好的相关性。眶上动脉可能是前额血压监测的更佳部位,因为它起源于颈内动脉,且在解剖学上更容易定位。本研究比较了通过示波法在眶上动脉和上臂测量的平均血压。
在全身麻醉下,对20名手术患者使用附着于前额的2.5×1厘米气囊,通过自粘垫记录眶上动脉的示波信号。使用Dinamap 1846血压监测仪同时测量手臂血压,并将所得数据与眶上动脉测量值进行比较。
前额和手臂的219对血压测量值之间的平均差异为3.8毫米汞柱。差异的标准差为7.4毫米汞柱。数据的线性回归方程为y = 0.98x + 3.25,估计标准误差为7.31毫米汞柱。两次测量之间的相关系数为0.82。
结果表明,通过示波法从眶上动脉和肱动脉测量的平均血压相互一致且相关性良好。眶上动脉应该是血压测量的一个良好替代部位。