Hashida S, Ishikawa S, Nakamoto H, Tanaka S, Kojima M, Ishikawa E
Department of Biochemistry, Medical College of Miyazaki, Japan.
J Clin Lab Anal. 1996;10(5):294-301. doi: 10.1002/(SICI)1098-2825(1996)10:5<294::AID-JCLA10>3.0.CO;2-S.
In a previously reported ultrasensitive enzyme immunoassay (immune complex transfer enzyme immunoassay) for antibody IgG to HIV-1, polystyrene beads in test tubes were handled with tweezers, and bound beta-D-galactosidase activity was measured with a fluorometer. The use of tweezers was causative of false-positivity by carryover, and testing many samples was difficult. Recently, these drawbacks have been minimized using microplates, a fluororeader, and modified polystyrene beads with sticks for easy handling. The modified polystyrene beads were transferred from wells to wells more quickly and easily without tweezers, eliminating false-positivity due to carryover. The fluorescence intensity for bound beta-D-galactosidase activity was quickly measured with a fluororeader. As a result, it became easy to test many samples with high reliability. However, modified polystyrene beads used were handmade, and the sensitivity was lower than the previous assay with test tubes. In the present study, the condition of the immune complex transfer enzyme immunoassay for antibody IgG to reverse transcriptase of HIV-1 was optimized using microplates, a fluororeader, and polystyrene solid phase with stick, which is commercially available. Testing many samples became easy with higher sensitivity.
在先前报道的用于检测抗HIV-1抗体IgG的超敏酶免疫测定法(免疫复合物转移酶免疫测定法)中,用镊子处理试管中的聚苯乙烯珠,并使用荧光计测量结合的β-D-半乳糖苷酶活性。使用镊子会因残留导致假阳性,并且难以检测许多样品。最近,使用微孔板、荧光读数器和带有便于操作的棒的改性聚苯乙烯珠,这些缺点已被最小化。改性聚苯乙烯珠无需镊子即可更快、更轻松地在孔之间转移,消除了因残留导致的假阳性。用荧光读数器快速测量结合的β-D-半乳糖苷酶活性的荧光强度。结果,以高可靠性检测许多样品变得容易。然而,使用的改性聚苯乙烯珠是手工制作的,并且灵敏度低于先前使用试管的测定法。在本研究中,使用微孔板、荧光读数器和市售的带棒聚苯乙烯固相优化了用于检测抗HIV-1逆转录酶抗体IgG的免疫复合物转移酶免疫测定条件。以更高的灵敏度轻松检测许多样品。