Röhrle H, Scholten R, Sollbach W, Ritter G, Grünert A
Arch Orthop Unfallchir. 1977 Jul 29;89(1):49-60. doi: 10.1007/BF00414825.
The mechanical force flow distribution produced by the hip joint endoprosthesis in the remaining, living femur and its deviation from that of the natural bone have a considerable influence on the quality of the bone-endoprosthesis compound and its duration of life. The loads are calculated for the natural bone as the reference system as well as the compound with conventional endoprostheses and a new type, the Tension-band-Endoprosthesis (R-G). For the calculation the finite element method is used by application of three-dimensional tetrahedron elements. It becomes apparent that in the case of the R-G-Endoprosthesis the induced stresses within the medial range of the remaining bone are far less unphysiological than in the case of the conventional endoprosthesis.
髋关节假体在剩余的活体股骨中产生的机械力流分布及其与天然骨的差异,对骨 - 假体复合物的质量及其使用寿命有相当大的影响。以天然骨作为参考系统计算载荷,同时也计算了与传统假体以及一种新型假体——张力带假体(R - G)组成的复合物的载荷。计算采用有限元方法,应用三维四面体单元。结果表明,与传统假体相比,R - G假体在剩余骨内侧范围内产生的诱导应力远没有那么不符合生理情况。