Multicenter comparison of truncated biphasic shocks and standard damped sine wave monophasic shocks for transthoracic ventricular defibrillation. Transthoracic Investigators.

作者信息

Bardy G H, Marchlinski F E, Sharma A D, Worley S J, Luceri R M, Yee R, Halperin B D, Fellows C L, Ahern T S, Chilson D A, Packer D L, Wilber D J, Mattioni T A, Reddy R, Kronmal R A, Lazzara R

机构信息

Department of Medicine, University of Washington (Seattle), USA.

出版信息

Circulation. 1996 Nov 15;94(10):2507-14. doi: 10.1161/01.cir.94.10.2507.

Abstract

BACKGROUND

The most important factor for improving out-of-hospital ventricular fibrillation survival rates is early defibrillation. This can be achieved if small, lightweight, inexpensive automatic external defibrillators are widely disseminated. Because automatic external defibrillator size and cost are directly affected by defibrillation waveform shape and because of the favorable experience with truncated biphasic waveforms in implantable cardioverter-defibrillators, we compared the efficacy of a truncated biphasic waveform with that of a standard damped sine monophasic waveform for transthoracic defibrillation.

METHODS AND RESULTS

The principal goal of this multicenter, prospective, randomized, blinded study was to compare the first-shock transthoracic defibrillation efficacy of a 130-J truncated biphasic waveform with that of a standard 200-J monophasic damped sine wave pulse using anterior thoracic pads in the course of implantable cardioverter-defibrillator testing. Pad-pad ECGs were also examined after transthoracic defibrillation. After the elimination of data for 24 patients who did not meet all protocol criteria, the results from 294 patients were analyzed. The 130-J truncated biphasic pulse and the 200-J damped sine wave monophasic pulse resulted in first-shock efficacy rates of 86% and 86%, respectively (P = .97). ST-segment levels measured 10 seconds after the shock in 151 patients in sinus rhythm were -0.26 +/- 1.58 and -1.86 +/- 1.93 mm for the 130- and 200-J shocks, respectively (P < .0001).

CONCLUSIONS

We found that 130-J biphasic truncated transthoracic shocks defibrillate as well as the 200-J monophasic damped sine wave shocks that are traditionally used in standard transthoracic defibrillators and result in fewer ECG abnormalities after the shock.