Dangas G, Mehran R, Duvvuri S, Ambrose J A, Sharma S K
Cardiovascular Institute, Mount Sinai Medical Center, New York City 10029, USA.
Cathet Cardiovasc Diagn. 1996 Oct;39(2):138-42. doi: 10.1002/ccd.1810390202.
Pseudoaneurysm (PA) formation is a well-documented complication of femoral arterial puncture in patients undergoing percutaneous cardiac procedures. Besides surgical repair, there has been success in obliterating the PAs by Duplex ultrasound guided compression (mostly those < 4.0 cm). We analyzed the use of prolonged femoral compression using a pneumatic compression system (FemoStop by USCI) in 10 patients who developed femoral PAs following an interventional cardiac procedure. All PAs were diagnosed by ultrasound and had unsuccessful duplex-guided compression. The size of the PA ranged from 2.8-4.0 cm in diameter. All patients had the FemoStop system placed over the PA with continuous pressure for 12-18 hr. PA obliteration was successful in 90% of cases as confirmed by ultrasound. The only complication observed was skin abrasion in two patients. Our results indicate that the FemoStop system is a simple and effective technique for treatment of femoral artery PAs.
假性动脉瘤(PA)形成是接受经皮心脏手术患者股动脉穿刺的一种有充分文献记载的并发症。除了手术修复外,通过双功超声引导压迫(主要是那些直径<4.0 cm的假性动脉瘤)成功消除了假性动脉瘤。我们分析了在10例介入性心脏手术后发生股动脉假性动脉瘤的患者中使用气动压迫系统(美国超声公司生产的FemoStop)进行长时间股动脉压迫的情况。所有假性动脉瘤均通过超声诊断,并且双功超声引导压迫未成功。假性动脉瘤的直径范围为2.8 - 4.0 cm。所有患者均将FemoStop系统置于假性动脉瘤上方并持续加压12 - 18小时。经超声确认,90%的病例假性动脉瘤消除成功。观察到的唯一并发症是两名患者出现皮肤擦伤。我们的结果表明,FemoStop系统是治疗股动脉假性动脉瘤的一种简单有效的技术。