Cyclic variation of Doppler power from whole blood under pulsatile flow.

作者信息

Wu S J, Shung K K

机构信息

Bioengineering Program, Pennsylvania State University, University Park 16802, USA.

出版信息

Ultrasound Med Biol. 1996;22(7):883-94. doi: 10.1016/0301-5629(96)00109-3.

Abstract

The echogenicity and Doppler power from whole blood under pulsatile flow have been found to vary during the flow cycle in previous studies both in vitro and in vivo. The present study was undertaken to better understand this phenomenon. Doppler power from whole blood under pulsatile flow was measured with a pulsed Doppler flowmeter as a function of the flow cycle, radial position and compliance of the vessel in a mock flow loop. It was found that the cyclic variation is more pronounced if the stroke rate is less than 56 beats/min and that the peak of the Doppler power from whole blood flowing near the center stream coincided with the peak of the flow velocity. However, it began to lead the velocity peak as the measurement site was moved away from the center stream. The lead increased as the radial distance was increased. The results also show that the compliance of the vessel can affect, to a certain extent, the magnitude of the cyclic variation. Results from intravascular Doppler measurements rule out the possibility that the cyclic variation is primarily due to the variation in attenuation caused by vessel wall during a flow cycle.