Physical and clinical evaluation of a 2,048 x 2,048-matrix image intensifier TV digital imaging system in bone radiography.

作者信息

Higashida Y, Baba Y, Hatemura M, Yoshida A, Takada T, Takahashi M

机构信息

Department of Radiological Technology, Kumamoto University College of Medical Science, Japan.

出版信息

Acad Radiol. 1996 Oct;3(10):842-8. doi: 10.1016/s1076-6332(96)80274-1.

Abstract

RATIONALE AND OBJECTIVES

To evaluate the potential utility of a 2,048 x 2,048-matrix image intensifier television digital radiography (DR) system versus a conventional screen-film (S-F) system for bone radiography.

METHODS

Basic imaging properties were evaluated including resolution properties, Wiener spectra, and detectabilities of low-contrast signals. DR images were obtained with the same exposure (iso-dose) or one-third the exposure (low-dose) used with the S-F system. The visibility of pathologic details of metastatic disease on bone radiographs of 27 patients was evaluated subjectively by six radiologists.

RESULTS

Resolution properties of the DR system were slightly superior to those of the S-F system at low-frequency range, but the S-F system showed considerably higher resolution properties at the high-frequency range. The noise levels for iso-dose DR were slightly greater than those for S-F imaging at low spatial frequency; however, low-dose digital radiographs showed higher noise levels. Visibility of details of diagnostic features on bone radiographs was similar with both systems, but the low-dose digital radiographs were slightly inferior.

CONCLUSION

High-resolution image intensifier television DR systems may be clinically useful for bone radiography.