An in-house enzyme-linked immunoabsorbant assay (ELISA) for SP-A was successfully developed using in-house polyclonal anti SP-A and a commercial polyclonal anti-rabbit immunoglobulin horseradish peroxidase conjugate system. The standard curve, generated by using 50 ng of SP-A to coat the plate and 1:500 dilution of polyclonal anti SP-A as a primary antibody, was linear for concentrations of SP-A ranging from 4 micrograms/l to 4000 micrograms/l and reproducible. Results of recovery study of SP-A from a known sample of tracheal aspirate ranged from 94%-114%. Intra- and inter-assay coefficients of variations were 2.7% and 5.6% respectively for a known sample of tracheal aspirate. Interference study showed that tracheal aspirate did not interfere with the assay. The assay developed was intended to be used for SP-A measurement in tracheal aspirates obtained from neonates with and without respiratory distress syndrome.