Earls J P, Rofsky N M, DeCorato D R, Krinsky G A, Weinreb J C
Department of Radiology, New York University Medical Center, NY 10016, USA.
Radiology. 1996 Dec;201(3):705-10. doi: 10.1148/radiology.201.3.8939219.
To determine the usefulness of a timing examination performed with 1 mL gadopentetate dimeglumine administered with a magnetic resonance (MR) power injector at single-dose three-dimensional (3D) MR aortography.
3D fast imaging with steady-state precession (repetition time [msec]/echo time [msec] = 5.0/2.0, with 50 degrees flip angle) was performed in 30 patients after administration of 0.1 mmol/kg gadopentetate dimeglumine. In 15 inpatients, a power injector was used and a timing examination was performed (experimental group). In 15 outpatients, manual injection was used and a timing examination was not performed (control group). In the timing examination, 1 mL gadopentetate dimeglumine was followed with 15 mL saline, both at 2 mL/sec, during axial turbo fast low-angle shot imaging (11/4.2/300 [inversion time ¿msec¿], with a 15 degrees flip angle); one image was obtained every 2 seconds for 60 seconds. Aortic signal-to-noise ratios (S/Ns) and contrast-to-noise ratios (C/Ns) were determined on images obtained in the two groups.
In the timing examinations, aortic S/N increased significantly (902% +/- 517 [P < .001]) and was sufficient to help determine the timing of arterial enhancement. Findings on experimental images were superior to those on control images (S/N, 29.8 vs 20.5 [P = .003]; C/N, 23.8 vs 12.8 [P < .001], respectively). The time to perform procedures and calculations at the timing examination added a mean of 4.5 minutes to each examination.
The combined use of a power injector and a timing examination improved S/Ns and C/Ns at breath-hold single-dose gadolinium-enhanced 3D MR aortography.
确定在单剂量三维磁共振血管造影中,使用磁共振(MR)高压注射器注射1 mL钆喷酸葡胺进行时间测定检查的效用。
对30例患者静脉注射0.1 mmol/kg钆喷酸葡胺后,采用三维稳态进动快速成像(重复时间[毫秒]/回波时间[毫秒]=5.0/2.0,翻转角50°)。15例住院患者使用高压注射器并进行时间测定检查(实验组)。15例门诊患者采用手动注射且未进行时间测定检查(对照组)。在时间测定检查中,以2 mL/秒的速度先后注射1 mL钆喷酸葡胺和15 mL生理盐水,同时进行轴位快速低角度激发成像(11/4.2/300[反转时间,毫秒],翻转角15°);每2秒采集一幅图像,共采集60秒。测定两组图像上的主动脉信噪比(S/N)和对比噪声比(C/N)。
在时间测定检查中,主动脉S/N显著升高(902%±517[P<.001]),足以帮助确定动脉强化的时间。实验组图像的表现优于对照组图像(S/N分别为29.8和20.5[P=.003];C/N分别为23.8和12.8[P<.001])。时间测定检查的操作和计算时间使每次检查平均增加4.5分钟。
在屏气单剂量钆增强三维磁共振血管造影中,联合使用高压注射器和时间测定检查可提高S/N和C/N。