股骨和胫骨X线片中髓内模板的准确性。

Accuracy of intramedullary templates in femoral and tibial radiographs.

作者信息

Krettek C, Blauth M, Miclau T, Rudolf J, Könemann B, Schandelmaier P

机构信息

Trauma Department, Hannover Medical School, Germany.

出版信息

J Bone Joint Surg Br. 1996 Nov;78(6):963-4. doi: 10.1302/0301-620x78b6.1284.

Abstract

The accuracy of templates used for the preoperative planning of the fixation of intramedullary fractures depends on radiological magnification. To study the accuracy of these templates, we randomly selected 100 femoral and 100 tibial radiographs taken after stabilisation by an intramedullary nail using a standard technique. We then compared the known nail length with the corresponding measurements on the radiographs. The mean magnification factor for the femur was 9% and for the tibia 7%; these differ considerably from the range of magnification of the manufacturers' templates (femur, 15% to 17%; tibia 10% to 15%). We conclude that templates are unreliable for the selection of implant length and that this should be done by intraoperative measurements.

摘要

用于髓内骨折固定术前规划的模板准确性取决于放射放大率。为研究这些模板的准确性,我们随机选取了100例股骨和100例胫骨的X线片,这些X线片是在采用标准技术通过髓内钉固定后拍摄的。然后我们将已知的钉长度与X线片上的相应测量值进行比较。股骨的平均放大系数为9%,胫骨为7%;这些值与制造商模板的放大率范围(股骨,15%至17%;胫骨,10%至15%)有很大差异。我们得出结论,模板在选择植入物长度方面不可靠,应通过术中测量来进行。