Design and comparison of wedge shaped midline block with rectangular block used in early stage cervical cancers.

作者信息

Avadhani J S, Deshpande D D, Pradhan A S, Viswanathan P S, Shrivastava S K, Dinshaw K A

机构信息

Department of Medical Physics, Tata Memorial Hospital, Mumbai, India.

出版信息

Strahlenther Onkol. 1996 Dec;172(12):676-80.

PMID:8972752
Abstract

PURPOSE

Design of an universal wedge shaped midline block for early stage cervical cancers to get uniform dose distribution around the target volume.

MATERIAL AND METHODS

The method of fabrication of wedge shaped midline block is discussed for treatment of early stage cervical cancers based on the configuration of 137Cs pellets used in intracavitary applications and along with external photon beams obtained by using 60Co teletherapy machine and 6 and 10 MV linear accelerators.

RESULTS

The dosimetric measurements are carried out to confirm the adequacy of thickness and shape of the block using radiation field analyzer for all the 3 energies.

DISCUSSION

The dose distribution comparison has been made with wedge shaped midline block and conventional rectangular shields to show its superiority in getting homogeneous dose distribution around target volume. The shift in 184 intracavitary applications is analyzed in actual clinical applications with respect to central axis of external beams and its comparison is made with both types of blocks. With the shift of 1 cm towards lateral from midline an overdose can occur upto 12 Gy at certain points with rectangular block compared to wedge shaped midline block.

CONCLUSION

Since the preparation of individual wedge shaped midline block for every application will be a difficult process for busy centers, the single wedge shaped midline block for each energy is more convenient to handle for busy centers instead of individualized compensating blocks for every intracavitary application.