[Quality assurance in radiology: quality control of radiographic cassettes].

作者信息

Ferretti P P, Sarti M, Messori P, Boni L, Seligardi P, Tassoni D, Cattini V, Piccagli V, Barani A, Bianchi C, Borasi G, Troiso A, Soliani Raschini C

机构信息

Servizio di Fisica Sanitaria, Arcispedale S. Maria Nuova, Azienda Ospedaliera di Reggio Emilia.

出版信息

Radiol Med. 1996 Sep;92(3):267-73.

PMID:8975314
Abstract

A "quality team" in radiology, whose members are the authors of this paper, has implemented a quality control program to test the cassettes with intensifying screen systems used in radiology departments. 149 systems-124 of them for general purpose radiology and 25 for mammography-were submitted to the following tests: visual inspection of radiographic cassettes and intensifying screens, screen-film contact, intensifying screen cleanliness and relative sensitivity of the intensifying screens. The results of each type of test are reported in detail in the paper, on a 3-point scale: good, sufficient and poor. The overall results of the quality control tests show 78% of general purpose radiology cassettes to qualify as good (69%) or sufficient (9%), while 22% were of poor quality. 88% of the mammographic cassettes qualified as good (76%) or sufficient (12%), while 12% were of poor quality. All tests were easy to perform and required limited resources. The necessary procedures to keep quality high over time are also reported. To conclude, the results obtained with our quality control program could be used as an effective tool to address and plan the turnover of the cassettes with intensifying screens which are usually used in diagnostic radiology practice.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验