Mochizuki T, Murase K, Tanaka H, Kondoh T, Hamamoto K, Tauxe W N
Department of Radiology, Ehime-Imabari Hospital, Japan.
J Nucl Med. 1997 Jan;38(1):53-7.
We evaluated ECG-gated SPECT (g-SPECT) in the measurement of absolute left ventricular (LV) volume by comparing it with left ventriculography (LVG) and with cine-MRI.
Projection data from 31 patients were acquired with a three-headed SPECT system in 12 min using a 64 x 64 matrix with 1.5 zoom (1 pixel = 4.27 mm). The R-R interval from simultaneously acquired ECG was divided into eight frames. The end-diastolic and end-systolic volumes (EDV; ESV) and LV mass were assessed by an area-length method with manual delineation of the epi- and endocardial LV borders using midventricular vertical and horizontal long-axis images. The stroke volume, LVEF and cardiac output (CO) were generated from the EDV, ESV and heart rate during the study. The g-SPECT LV values were compared with those of LVG (25 patients) and cine-MRI (18 patients).
The g-SPECT values correlated well with those from LVG (r = 0.83 to 0.92; p < 0.001) and cine-MRI (r = 0.76 to 0.99; p < 0.001). The g-SPECT technique provides an assessment of LV volumes (EDV, ESV, stroke volume, LVEF, CO, LV mass).
Despite potential problems that may cause inaccuracy and require improvements such as an accurate and reproducible automatic edge detection algorithm, g-SPECT has clinical utility in assessing global LV volumes and function.
我们通过将心电图门控单光子发射计算机断层扫描(g-SPECT)与左心室造影(LVG)及电影磁共振成像(cine-MRI)相比较,评估其在测量左心室(LV)绝对容积方面的性能。
使用三头SPECT系统,以64×64矩阵、1.5倍放大率(1像素 = 4.27毫米)在12分钟内采集31例患者的投影数据。将同时采集的心电图的R-R间期分为8帧。使用心室中部垂直和水平长轴图像,通过面积-长度法手动勾勒LV的心外膜和心内膜边界,评估舒张末期和收缩末期容积(EDV;ESV)以及LV质量。在研究期间,根据EDV、ESV和心率计算每搏输出量、左心室射血分数(LVEF)和心输出量(CO)。将g-SPECT的LV值与LVG(25例患者)和cine-MRI(18例患者)的LV值进行比较。
g-SPECT值与LVG(r = 0.83至0.92;p < 0.001)和cine-MRI(r = 0.76至0.99;p < 0.001)的值具有良好的相关性。g-SPECT技术可对LV容积(EDV、ESV、每搏输出量、LVEF、CO、LV质量)进行评估。
尽管g-SPECT可能存在一些导致不准确且需要改进的潜在问题,如准确且可重复的自动边缘检测算法,但它在评估整体LV容积和功能方面具有临床实用性。