Chauveau N, Ayeva B, Rigaud B, Morucci J P
INSERM U305, Hôtel Dieu, Toulouse, France.
Physiol Meas. 1996 Nov;17 Suppl 4A:A7-13. doi: 10.1088/0967-3334/17/4a/003.
A multifrequency (1 kHz-1 MHz) serial electrical impedance tomography (EIT) system has been developed. It is based on 16 active electrodes and can be extended up to 32. Each active electrode can be programmed for current driving and for measuring either the injected current or the voltage difference between adjacent electrodes, and includes calibration facilities. Real and imaginary parts of the impedance are obtained by applying a parametric identification method (extended Prony), but other techniques are easily adaptable. Image reconstruction is carried out using the Sheffield filtered back-projection algorithm. Characteristic frequency images are under development and should be of great interest to distinguish between normal and tumorous tissues.
已开发出一种多频率(1千赫至1兆赫)串行电阻抗断层成像(EIT)系统。它基于16个有源电极,可扩展至32个。每个有源电极可进行编程,用于电流驱动以及测量注入电流或相邻电极之间的电压差,并包括校准功能。通过应用参数识别方法(扩展Prony法)获得阻抗的实部和虚部,但其他技术也易于适配。使用谢菲尔德滤波反投影算法进行图像重建。特征频率图像正在开发中,对于区分正常组织和肿瘤组织应该会非常有用。