使用基于胶片的成像和计算机断层扫描对骨整合植入物进行剂量测定及成本分析。

Dosimetry and cost of imaging osseointegrated implants with film-based and computed tomography.

作者信息

Scaf G, Lurie A G, Mosier K M, Kantor M L, Ramsby G R, Freedman M L

机构信息

Faculdade de Odontologia de Araraquara-UNESP, School of Dental Medicine, Barzil.

出版信息

Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1997 Jan;83(1):41-8. doi: 10.1016/s1079-2104(97)90089-5.

Abstract

Thermoluminescent dosimeters were used to measure radiation doses at craniofacial sites in a tissue-equivalent phantom during film-based multidirectional tomography with the Tomax Ultrascan (Incubation Industries, Ivyland, Pa.) and during computed tomography with the Elscint Excel 2400 (Elscint Corp., Tel Aviv, Israel). Mean absorbed doses for presurgical mandibular and maxillary canine and molar implant assessments were converted to equivalent doses, which were then multiplied by published weighting factors and summed to give effective doses. The computed tomography device consistently delivered higher doses than the Tomax Ultrascan to all anatomic locations; the differences were most pronounced when only one or two implant sites were evaluated. The reasons for the dose disparities are considered both anatomically and procedurally. A survey of examination cost revealed film-based multidirectional tomography to be less expensive than computed tomography.

摘要

在使用Tomax Ultrascan(Incubation Industries,Ivyland,宾夕法尼亚州)进行基于胶片的多向断层扫描以及使用Elscint Excel 2400(Elscint公司,特拉维夫,以色列)进行计算机断层扫描期间,使用热释光剂量计测量组织等效体模中颅面部部位的辐射剂量。将术前下颌和上颌尖牙及磨牙种植体评估的平均吸收剂量转换为当量剂量,然后乘以已公布的权重因子并求和得出有效剂量。计算机断层扫描设备在所有解剖位置的剂量始终高于Tomax Ultrascan;当仅评估一两个种植部位时,差异最为明显。从解剖学和程序方面考虑了剂量差异的原因。一项检查成本调查显示,基于胶片的多向断层扫描比计算机断层扫描成本更低。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验