Vogl T J, König O, Böttcher H, Felix R
Strahlenklinik und Poliklinik, Klinikum Rudolf Virchow, Humboldt-Universität zu Berlin.
Rofo. 1996 Dec;165(6):568-73. doi: 10.1055/s-2007-1015813.
To evaluate parameters and characteristics of excimer-laser-angioplasty on ex-vivo vessel models.
450 tissue ablations were performed on pig aortas, human arteries and fibrin-fixed plaque material. A pulsed 308nm-XeCl-laser delivering 120 ns pulses was tested. Ablations were analysed depending on frequency (10-40 Hz), energy density (30-70 ml/mm2), catheter-material-angle (0-90 degrees) and object-catheter-distance (0-40 mm).
Vessel recanalisation speeds of up to 0.5 mm/s were achieved on fibrosed to slightly calcified human vessel models. In perforation tests on normal to slightly fibrosed human vessel walls and vessels of pigs ablation speeds of 0.01-0.5 mm/s were measured. Perforations on calcified vessels were successful in 50% of cases after two minutes application time. Vessel perforations were not detected without the support of mechanical power. At catheter-material angles < or = 30 degrees a minor damage of the intima could be found. There was no evidence of necrosis or coagulation in our studies.
In-vitro tests of human plaque material yield reproducible vessel recanalisation distances of up to 25 mm at a variable speed from 0.1-0.2 mm/s.
评估准分子激光血管成形术在离体血管模型上的参数和特性。
在猪主动脉、人体动脉和纤维蛋白固定的斑块材料上进行了450次组织消融。测试了一种发射120纳秒脉冲的脉冲308纳米氙氯激光器。根据频率(10 - 40赫兹)、能量密度(30 - 70毫升/平方毫米)、导管-材料角度(0 - 90度)和物体-导管距离(0 - 40毫米)对消融进行分析。
在纤维化至轻度钙化的人体血管模型上实现了高达0.5毫米/秒的血管再通速度。在对正常至轻度纤维化的人体血管壁和猪血管进行的穿孔测试中,测量到的消融速度为0.01 - 0.5毫米/秒。在钙化血管上,施加两分钟后,50%的病例穿孔成功。没有机械动力支持时未检测到血管穿孔。在导管-材料角度≤30度时,可发现内膜有轻微损伤。在我们的研究中没有坏死或凝血的证据。
对人体斑块材料的体外测试可产生可重复的血管再通距离,最高可达25毫米,速度在0.1 - 0.2毫米/秒之间变化。