使用MCNP研究计数几何对¹²⁵I监测的影响。

Examination of the effect of counting geometry on 125I monitoring using MCNP.

作者信息

Kramer G H, Yiu S

机构信息

Radiation Protection Bureau, Ottawa, Ontario, Canada.

出版信息

Health Phys. 1997 Mar;72(3):465-70. doi: 10.1097/00004032-199703000-00015.

Abstract

This paper theoretically investigates how the counting efficiencies of three selected detector (NaI) sizes are affected by neck-detector distance, detector misplacement, thickness of overlaying tissue, and size of the thyroid gland. The detector sizes were small, 2.54 cm diameter and 0.2 cm crystal thickness; medium, 7.62 cm diameter and 0.2 cm crystal thickness; large 30.48 cm diameter and 0.2 cm crystal thickness. The evaluations were determined using a Monte Carlo technique. The simulations show that the large detector minimizes the uncertainties for all problems except the thickness of overlaying tissue.

摘要

本文从理论上研究了三种选定尺寸的探测器(碘化钠)的计数效率如何受到颈部与探测器距离、探测器位置偏移、覆盖组织厚度以及甲状腺大小的影响。探测器尺寸分别为小尺寸,直径2.54厘米,晶体厚度0.2厘米;中尺寸,直径7.62厘米,晶体厚度0.2厘米;大尺寸,直径30.48厘米,晶体厚度0.2厘米。评估采用蒙特卡罗技术进行。模拟结果表明,除覆盖组织厚度外,大尺寸探测器能将所有问题的不确定性降至最低。