Suppr超能文献

眼电图比率及其对视网膜功能的评估。

The EOG ratio and its evaluation of retinal function.

作者信息

Afanador A J

出版信息

J Am Optom Assoc. 1977 Sep;48(9):1149-56.

PMID:903557
Abstract

The evaluation of 116 normal ratios with a 10 ft. lambert ganzfield illumination confirmed that ratios over 200% can be considered as normal. Individual variation led to the proposed classification of 175% to 200% as being low or suspicious. Ratios below 175% are abnormal and are useful in establishing the diagnosis of pathology.

摘要

在10英尺朗伯全视野照明条件下对116个正常比率进行评估,结果证实比率超过200%可视为正常。个体差异导致将175%至200%的比率分类为低或可疑。低于175%的比率为异常,有助于病理诊断的确立。

相似文献

8
[Fluorescein angiography and retinal pigmentary epithelium].
Ann Ocul (Paris). 1971 Nov;204(11):1167-84.
9
Retinitis pigmentosa and allied retinal diseases: electrophysiologic findings.
Trans Sect Ophthalmol Am Acad Ophthalmol Otolaryngol. 1976 Jul-Aug;81(4 Pt 1):OP659-666.
10
[Electrooculography].
Fortschr Med. 1972 Nov 16;90(32):1175-80.