The accuracy of a new system for estimating organ volume using ultrasound.

作者信息

Hughes S W, D'Arcy T J, Maxwell D J, Saunders J E, Chinn S, Sheppard R J

机构信息

Department of Medical Physics, Guy's and St Thomas' Hospital Trust, London, UK.

出版信息

Physiol Meas. 1997 Feb;18(1):73-84. doi: 10.1088/0967-3334/18/1/006.

Abstract

A new system is described for estimating volume from a series of multiplanar 2D ultrasound images. Ultrasound images are captured using a personal computer video digitizing card and an electromagnetic localization system is used to record the pose of the ultrasound images. The accuracy of the system was assessed by scanning four groups of ten cadaveric kidneys on four different ultrasound machines. Scan image planes were oriented either radially, in parallel or slanted at 30 degrees to the vertical. The cross-sectional images of the kidneys were traced using a mouse and the outline points transformed to 3D space using the Fastrak position and orientation data. Points on adjacent region of interest outlines were connected to form a triangle mesh and the volume of the kidneys estimated using the ellipsoid, planimetry, tetrahedral and ray tracing methods. There was little difference between the results for the different scan techniques or volume estimation alogorithms, although, perhaps as expected, the ellipsoid results were the least precise. For radial scanning and ray tracing, the mean and standard deviation of the percentage errors for the four different machines were as follows: Hitachi EUB-240, -3.0 +/- 2.7%; Tosbee RM3, -0.1 +/- 2.3%; Hitachi EUB-415, 0.2 +/- 2.3%; Acuson, 2.7 +/- 2.3%.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验