[Nepross C - 02透析器的临床经验]

[Clinical experiences with Nepross C-02 dialyzer].

作者信息

Balcke P, Schmidt P, Zazgornik J, Kopsa H, Pils P

出版信息

Acta Med Austriaca. 1977;4(2):47-9.

PMID:906781
Abstract

During 50 haemodialyses procedures with the Nephross C-02 coil dialyzer in a RSP-system clearance, effectivity and ultrafiltration were investigated. With a blood flow of 200 ml per minute clearance for urea, creatinine and uric acid were 88.3, 76.1 and 69.9 ml per minute respectively. Considering these results, our findings 2 hours after starting dialysis must be taken into account that clearance values measured in the recirculation-single pass-system (RSP) are about 30% lower than the results with the use of the single pass-system. The maximal ultrafiltration rate was 600 ml per hour. Overhydration therefore, is the main indication for the use of the Nephross C-02 coil dialyzer.

摘要

在使用Nephross C-02盘管式透析器进行的50次血液透析过程中,在RSP系统清除率、有效性和超滤方面进行了研究。血液流速为每分钟200毫升时,尿素、肌酐和尿酸的清除率分别为每分钟88.3毫升、76.1毫升和69.9毫升。考虑到这些结果,在开始透析2小时后我们的发现必须考虑到,在再循环单通道系统(RSP)中测得的清除率值比使用单通道系统的结果低约30%。最大超滤率为每小时600毫升。因此,水合过多是使用Nephross C-02盘管式透析器的主要指征。