Thompson J, Haven G
Am J Clin Pathol. 1977 Oct;68(4):474-80. doi: 10.1093/ajcp/68.4.474.
A radioimmunoassay using a commercially available antiserum was evaluated for measurement of serum unconjugated estriol in pregnancy. The evaluation showed an inter-assay variance of 12.1%, intra-assay variance of 6.8%, sensitivity of 0.2-0.6 ng/ml (0.7-2.1 nmol/l), and average recovery of 85.3%. The assay is specific for unconjugated estriol, showing less than 1% cross-reactivity with estriol-3-sulfate and estriol-16-glucuronide. Normal limits were established from 7 to 40 weeks' gestation using 175 serum samples. No diurnal variation could be demonstrated at 8 a.m. and 3 p.m. Eighty-nine serum specimens and 82 urine specimens obtained from 18 high-risk pregnancies were within normal limits except in cases of intruterine fetal death, pre-eclampsia, and suspected placental sulfatase deficiency. Serial urinary estriol levels fluctuated as much as 75%, while serial serum samples varied by only 30%.
对一种使用市售抗血清的放射免疫分析法进行了评估,以测定孕期血清中未结合雌三醇的含量。评估结果显示,批间变异系数为12.1%,批内变异系数为6.8%,灵敏度为0.2 - 0.6 ng/ml(0.7 - 2.1 nmol/l),平均回收率为85.3%。该分析法对未结合雌三醇具有特异性,与雌三醇 - 3 - 硫酸盐和雌三醇 - 16 - 葡萄糖醛酸的交叉反应率小于1%。利用175份血清样本确定了孕7至40周的正常范围。上午8点和下午3点未发现昼夜变化。从18例高危妊娠中获取的89份血清标本和82份尿液标本,除了宫内胎儿死亡、先兆子痫和疑似胎盘硫酸酯酶缺乏的病例外,均在正常范围内。系列尿雌三醇水平波动高达75%,而系列血清样本仅变化30%。