Suppr超能文献

用于基于框架的立体定向手术的磁共振图像失真校正方法及初步结果。

Method for correcting magnetic resonance image distortion for frame-based stereotactic surgery, with preliminary results.

作者信息

Sumanaweera T S, Adler J R, Glover G H, Hemler P F, van den Elsen P A, Martin D, Napel S

机构信息

Department of Radiology, Stanford University School of Medicine, California 94305-5488, USA.

出版信息

J Image Guid Surg. 1995;1(3):151-7. doi: 10.1002/(SICI)1522-712X(1995)1:3<151::AID-IGS4>3.0.CO;2-9.

Abstract

We previously described a technique for correcting patient-specific magnetic field inhomogeneity spatial distortion in magnetic resonance images (MRI), which was not applicable to patients fitted with MRI-compatible stereotactic fiducial frames. Here we describe an improvement to the technique that permits application for these patients. Measurements with a cadaver head show that this method achieves MRI stereotactic localization accuracy of 1 mm.

摘要

我们之前描述了一种校正磁共振成像(MRI)中患者特异性磁场不均匀性空间畸变的技术,该技术不适用于配备MRI兼容立体定向基准框架的患者。在此,我们描述了该技术的一种改进方法,使其能够应用于这些患者。对一具尸体头部的测量表明,该方法实现了1毫米的MRI立体定向定位精度。