Yu Z, Westerlund D, Boos K S
Uppsala University Biomedical Centre, Sweden.
J Chromatogr B Biomed Sci Appl. 1997 Feb 21;689(2):379-86. doi: 10.1016/s0378-4347(96)00357-x.
A simple, sensitive and fully automated column-switching system by direct injection of plasma samples for determination of methotrexate and its metabolite 7-hydroxymethotrexate was developed. The system utilized a C8 alkyl-diol silica precolumn coupled with a LiChrospher RP-18 analytical column, followed by a photoreactor and fluorimetric detection. The photo-oxidative irradiation was accomplished at UV 254 nm in the presence of 0.1% hydrogen peroxide in the eluent. Studies showed that the fluorimetric response was influenced by the reaction time, the degree of the reactor's transparency and the choice of the working wavelengths. By optimizing the content of acetonitrile in the eluent, methotrexate can be separated from 7-hydroxymethotrexate completely. The method validation revealed quantitative recoveries (> or = 94%) with coefficients of variation < or = 4.4%. The limits of detection and quantitation for determination of methotrexate were 0.20 and 0.36 ng, respectively, corresponding to 2.0 and 3.6 ng/ml for an injection volume of 100 microliters. It was possible to enhance the sensitivity further by injecting larger plasma volumes, up to 500 microliters.
建立了一种通过直接进样血浆样品来测定甲氨蝶呤及其代谢物7-羟基甲氨蝶呤的简单、灵敏且全自动的柱切换系统。该系统采用C8烷基二醇硅胶预柱与LiChrospher RP - 18分析柱联用,随后连接光反应器和荧光检测。光氧化照射在254 nm紫外光下,于洗脱液中存在0.1%过氧化氢的条件下完成。研究表明,荧光响应受反应时间、反应器透明度以及工作波长选择的影响。通过优化洗脱液中乙腈的含量,可将甲氨蝶呤与7-羟基甲氨蝶呤完全分离。方法验证显示定量回收率(≥94%),变异系数≤4.4%。测定甲氨蝶呤的检测限和定量限分别为0.20 ng和0.36 ng,对于100微升进样体积,分别相当于2.0 ng/ml和3.6 ng/ml。通过进样更大体积的血浆(高达500微升)有可能进一步提高灵敏度。