脊髓损伤患者肩旋转肌力量的等速测力和手持测力计测量评估

Assessment of isokinetic and hand-held dynamometer measures of shoulder rotator strength among individuals with spinal cord injury.

作者信息

May L A, Burnham R S, Steadward R D

机构信息

Faculty of Rehabilitation Medicine, University of Alberta, Edmonton, Canada.

出版信息

Arch Phys Med Rehabil. 1997 Mar;78(3):251-5. doi: 10.1016/s0003-9993(97)90029-0.

Abstract

OBJECTIVE

To determine the intrarater reliability of the hand-held dynamometer (HHD) for the measurement of shoulder rotation and to examine the relationship between the peak force values of the HHD and the isokinetic dynamometer.

PARTICIPANTS

Twenty-five spinal cord injured individuals, 12 persons with paraplegia and 13 persons with tetraplegia between the ages of 18 and 42 years, were recruited from the community.

SETTING

Private practice clinic.

PROCEDURES

All participants were tested with the HHD by one examiner to determine reliability. On the same day, these participants underwent isokinetic testing to determine the relationship of the HHD and Cybex dynamometer measurements.

RESULTS

The intraclass correlation coefficients for the intrarater reliability ranged from .89 to .96. The Pearson product moment correlation was used to analyze the relationship between the two devices. All coefficients (.52 to .88) were statistically significant (p < .01); however, separate analyses for the persons with paraplegia and tetraplegia differed considerably. There was substantial variability of the isokinetic strength values at the lower levels of isometric strength.

CONCLUSIONS

The HHD can be used reliably to measure shoulder rotation in paraplegic and tetraplegic spinal cord injured individuals. Although it appears that the relationship between HHD and isokinetic measurement is poor for the participants with tetraplegia, the variability of the isokinetic scores indicates that this observation may be a function of the method of isokinetic measurements. Further study with a modified isokinetic testing protocol is needed to clarify the results of the participants with tetraplegia.

摘要

目的

确定手持测力计(HHD)测量肩部旋转的评定者内信度,并检验HHD峰值力值与等速测力计之间的关系。

参与者

从社区招募了25名脊髓损伤个体,年龄在18至42岁之间,其中12人患有截瘫,13人患有四肢瘫。

地点

私人诊所。

程序

由一名检查者使用HHD对所有参与者进行测试以确定信度。在同一天,这些参与者接受等速测试以确定HHD与Cybex测力计测量结果之间的关系。

结果

评定者内信度的组内相关系数范围为0.89至0.96。使用Pearson积差相关分析两种设备之间的关系。所有系数(0.52至0.88)均具有统计学意义(p <.01);然而,对截瘫和四肢瘫患者的单独分析有很大差异。在较低等长力量水平下,等速力量值存在很大变异性。

结论

HHD可可靠地用于测量截瘫和四肢瘫脊髓损伤个体的肩部旋转。尽管对于四肢瘫患者,HHD与等速测量之间的关系似乎较差,但等速评分的变异性表明这一观察结果可能是等速测量方法的作用。需要采用改良的等速测试方案进行进一步研究,以阐明四肢瘫患者的结果。