Hong M K, Beyar R, Kornowski R, Tio F O, Bramwell O, Leon M B
Department of Internal Medicine (Cardiology Division), Washington Hospital Center, Washington, DC, USA.
Coron Artery Dis. 1997 Jan;8(1):45-8. doi: 10.1097/00019501-199701000-00007.
There is a paucity of experimental data regarding self-expanding stents. This study evaluated the acute and chronic effects of CardioCoil, a self-expanding nitinol coil stent, in porcine coronary arteries.
Twenty-three self-expanding nitinol stents were implanted without associated balloon angioplasty in normal coronary arteries of 12 pigs, which were serially sacrificed up to 6 months. Angiographic and histologic analyses were performed to evaluate the deployment characteristics, patency rates, neointimal response, and unique features of the self-expanding nature of the CardioCoil stent.
All stents were successfully deployed and remained patent acutely. Three undersized stents migrated proximally and there was one episode of subacute thrombosis in an oversized stent. The remaining stents were patent throughout the survival period and neointimal responses were favorable for up to 6 months (all mean neointima < 200 microns up to 6 months). There was evidence of continuing stent expansion over time (stent diameter 2.85 +/- 0.78 mm immediately after deployment and 3.24 +/- 0.97 mm at follow-up) and the majority of stent struts were in the adventitia by 6 months. Re-endothelization occurred starting one week after implantation and was complete by 8 weeks.
This study shows that the CardioCoil self-expanding nitinol coil stent, is associated with favorable deployment characteristics and patency rates, although appropriate sizing is more crucial than with balloon-expandable stents. More importantly, there appears to be a "dissociation' between the deep vessel wall injury by the chronic strut expansion process and the neointimal reaction, unlike balloon-expandable stents.
关于自膨胀支架的实验数据较少。本研究评估了自膨胀镍钛合金线圈支架CardioCoil在猪冠状动脉中的急性和慢性影响。
将23个自膨胀镍钛合金支架在未进行相关球囊血管成形术的情况下植入12头猪的正常冠状动脉中,这些猪在长达6个月的时间里被依次处死。进行血管造影和组织学分析,以评估CardioCoil支架的展开特性、通畅率、新生内膜反应以及自膨胀特性的独特特征。
所有支架均成功展开,急性时保持通畅。3个尺寸过小的支架向近端迁移,1个尺寸过大的支架发生了1次亚急性血栓形成。其余支架在整个存活期内保持通畅,新生内膜反应在长达6个月时良好(6个月时所有平均新生内膜<200微米)。有证据表明支架随时间持续扩张(展开后即刻支架直径为2.85±0.78毫米,随访时为3.24±0.97毫米),到6个月时大多数支架支柱位于外膜。植入后1周开始重新内皮化,8周时完成。
本研究表明,CardioCoil自膨胀镍钛合金线圈支架具有良好的展开特性和通畅率,尽管合适的尺寸选择比球囊扩张支架更为关键。更重要的是,与球囊扩张支架不同,慢性支柱扩张过程对血管壁深层的损伤与新生内膜反应之间似乎存在“分离”。