Mattke S, Kramm B
Ludwig-Maximilians-Universität München, Medizinische Klinik I, Munich, Germany.
Pacing Clin Electrophysiol. 1997 Apr;20(4 Pt 1):993-4. doi: 10.1111/j.1540-8159.1997.tb05507.x.
We report on inadequate shocks delivered by an ICD for VT converting spontaneously to AF with slow ventricular response. This represents a correct behavior of the reconfirmation algorithm, as the detection of intervals longer than the fibrillation detection interval and shorter than the bradycardia escape interval is required to inhibit the shock.
我们报告了一例植入式心律转复除颤器(ICD)在室性心动过速(VT)自发转为伴有缓慢心室反应的心房颤动(AF)时发放不适当电击的情况。这体现了再确认算法的正确行为,因为需要检测到长于颤动检测间期且短于心动过缓逸搏间期的间期来抑制电击发放。