A precision repeat localization head frame for fractionated stereotactic radiotherapy.

作者信息

Scott T W, Beach J L, Mendiondo O A

机构信息

Lexington Radiation Therapy Center, KY 40504, USA.

出版信息

Med Dosim. 1997 Spring;22(1):5-8. doi: 10.1016/s0958-3947(96)00148-3.

Abstract

An immobilization device was constructed for Fractionated Stereotactic Radiotherapy (FSRT) based on registration of the teeth and facial bones in a single thermoplastic mask system, along with a custom hardened foam pillow for posterior head immobilization relative to the mask. The unit interfaces mechanically with all of our current radiosurgery equipment and can be used with any standard stereotactic planning system. After initial trials to design a reproducible radiographic localization test, we performed a series of daily AP and Lateral port films on 3 patients over five isocenters. Seventy-nine films were reviewed and the maximum deviation in anatomical projection in both sagittal and coronal planes was less than 2 mm, with over 60% of films showing no distinguishable deviations from initial port films. Ninety-three percent of the test films showed a repositioning accuracy of less than 1 mm for all tested structures. We have developed an accurate, non-invasive means of repeat head immobilization that, when properly constructed, can facilitate precise fractionated stereotactic radiation therapy with patient comfort, ease of construction and long term stability.

摘要

基于单个热塑性面罩系统中牙齿和面部骨骼的配准,构建了一种用于分次立体定向放射治疗(FSRT)的固定装置,以及一个定制的硬化泡沫枕头,用于相对于面罩对头后部进行固定。该装置与我们现有的所有放射外科设备实现机械对接,可与任何标准的立体定向治疗计划系统配合使用。在进行初步试验以设计可重复的放射成像定位测试后,我们对3名患者在5个等中心进行了一系列每日前后位(AP)和侧位端口片拍摄。共审查了79张片子,矢状面和冠状面的解剖投影最大偏差均小于2毫米,超过60%的片子与初始端口片相比无明显偏差。93%的测试片子显示,所有测试结构的重新定位精度均小于1毫米。我们开发了一种准确、无创的重复头部固定方法,该方法构建得当,可为精确的分次立体定向放射治疗提供便利,同时确保患者舒适度高、易于构建且具有长期稳定性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验