Both multicrystal and single-crystal detectors have been in use for more than 25 years for measurement of ejection fraction by analysis of images collected during the first-pass transit of radionuclides through the heart. Originally, multicrystal cameras were preferred, because they provided higher count rates than Anger cameras; however, over the years improvements in count rate capability and collimator design have enabled Anger cameras to perform equally well. This has become an important issue now that readily available 99mTc agents, such as sestamibi, enable evaluation of both myocardial function and perfusion from a single injection. The technical abilities of a particular camera determine which acquisition protocols are most likely to provide clinically useable images for the widest spectrum of patients. Electrocardiographic-gated list mode collection is highly desirable for first-pass imaging, providing the greatest flexibility of data review, rebinning, and analysis. Attention to quality control issues of data characterization and processing is important to ensure accuracy and precision of all measurements. Accurate determinations of ejection fraction of the left ventricle are possible routinely and, under favorable circumstances, of the right ventricle as well.
多晶体探测器和单晶探测器用于通过分析放射性核素首次通过心脏时采集的图像来测量射血分数已有25年以上的时间。最初,多晶体相机更受青睐,因为它们比安格尔相机提供更高的计数率;然而,多年来计数率能力和准直器设计的改进使安格尔相机能够表现得同样出色。鉴于诸如 sestamibi 等易于获得的 99mTc 试剂能够通过单次注射评估心肌功能和灌注,这已成为一个重要问题。特定相机的技术能力决定了哪些采集协议最有可能为最广泛的患者群体提供临床可用图像。心电图门控列表模式采集对于首次通过成像非常理想,它提供了最大的数据审查、重新分箱和分析灵活性。关注数据表征和处理的质量控制问题对于确保所有测量的准确性和精确性很重要。常规情况下可以准确测定左心室射血分数,在有利条件下也可以测定右心室射血分数。