Palmon L U, Foshager M C, Parantainen H, Everson L I, Cunningham B
Department of Radiology, University of Minnesota Hospital and Clinic, Minneapolis, MN 55455, USA.
AJR Am J Roentgenol. 1997 Jun;168(6):1595-8. doi: 10.2214/ajr.168.6.9168734.
During sonographic evaluation of silicone breast implants for possible rupture, we have frequently encountered several patterns of linear echoes within the implants. To our knowledge, the significance of this finding has not been established in the literature. The purpose of this study was to determine whether internal echoes are significant in predicting implant rupture.
Thirty-three patients with 64 silicone implants were prospectively entered into a study that included gray-scale sonography of the implants and subsequent surgical removal. Echo patterns within the implants were retrospectively evaluated on hard-copy films and compared with the integrity of the implant at surgery.
Three categories of internal echo patterns were identified: "thick linear echoes." "thin linear echoes," and "commas." One or more of these echo patterns were seen in 57 (89%) of the 64 implants. Thick linear echoes were seen in 23 (36%) of the 64 implants, thin linear echoes were seen in 33 (52%) of the 64 implants, and commas were seen in 47 (73%) of the 64 implants. All echo patterns were seen in intact and ruptured implants with nearly equal frequency. We found no statistical significance for any echo pattern in predicting whether an implant was ruptured or intact. Of the 64 implants, four were entirely free of internal echoes. All four implants were intact.
A variety of linear echoes can be seen in most silicone breast implants on gray-scale sonography. The presence or absence of linear echoes is not useful in predicting implant rupture. Complete absence of internal echoes, while highly predictive of an intact implant, is infrequently seen.
在对硅胶乳房植入物进行超声评估以判断是否可能破裂时,我们经常在植入物内遇到几种线性回声模式。据我们所知,这一发现的意义在文献中尚未得到证实。本研究的目的是确定内部回声在预测植入物破裂方面是否具有重要意义。
前瞻性纳入33例患者的64个硅胶植入物,进行包括植入物灰阶超声检查及随后手术取出的研究。回顾性地在硬拷贝胶片上评估植入物内的回声模式,并与手术时植入物的完整性进行比较。
识别出三类内部回声模式:“厚线性回声”、“薄线性回声”和“逗号样回声”。64个植入物中有57个(89%)出现了一种或多种这些回声模式。64个植入物中有23个(36%)出现厚线性回声,33个(52%)出现薄线性回声,47个(73%)出现逗号样回声。完整和破裂的植入物中出现所有回声模式的频率几乎相等。我们发现任何一种回声模式在预测植入物是否破裂或完整方面均无统计学意义。64个植入物中,有4个完全没有内部回声。这4个植入物均完整。
在大多数硅胶乳房植入物的灰阶超声检查中可看到多种线性回声。线性回声的有无对预测植入物破裂并无帮助。完全没有内部回声虽然高度提示植入物完整,但很少见。