Bach M
Univ.-Augenklinik, Killianstr, Freiburg, Germany.
Spat Vis. 1997;10(4):485-9. doi: 10.1163/156856897x00429.
Accurate luminance calibration is an important issue for stimuli in vision research. Raster-scan cathode-ray tubes present a rapidly scanning spot with a luminance of the order of 10,000 cd m-2 (the actual value depends on video timing and phosphor persistence). With little spatial integration this may result in overloading of the front stage of a photometer. More importantly, even if averaged over hundreds of milliseconds, the pulsed nature of the luminance signal will markedly reduce accuracy of the luminance measurement. A frame-synchronised photometer is described to increase measurement accuracy whilst still rapidly acquiring the result.
精确的亮度校准是视觉研究中刺激源的一个重要问题。光栅扫描阴极射线管呈现出一个快速扫描的光斑,其亮度约为10000 cd m-2(实际值取决于视频定时和磷光体余辉)。由于空间积分很少,这可能会导致光度计前端过载。更重要的是,即使在数百毫秒内进行平均,亮度信号的脉冲性质也会显著降低亮度测量的准确性。本文描述了一种帧同步光度计,以提高测量精度,同时仍能快速获得结果。