Methods for comparing the performance of different gamma cameras.

作者信息

Goodwin P N, Himelstein E

出版信息

Semin Nucl Med. 1977 Oct;7(4):299-304. doi: 10.1016/s0001-2998(77)80038-x.

DOI:10.1016/s0001-2998(77)80038-x
PMID:918671
Abstract

Meaningful comparisons of different gamma cameras require the acquisition of numerical data characterizing each of the characteristics of interest. These include resolution, contrast, sensitivity, uniformity, dead time (and the resulting percent loss), and the maximum counting rate. Resolution is best determined by using a line source to measure the linespread function, from which the modulation transfer function can be calculated. Bar patterns are useful in showing the effect on resolution of different collimators and different source distances. Sensitivity comparisons depend on collimator choice, and require counting the same object under identical conditions. Accurate comparisons of dead times and maximum counting rates require extreme care that measurements are taken under identical conditions of scatter, source volume, window setting, etc. Comparisons of the intrinsic dead times should be made with the collimator off, while actual counting losses should be determined using the collimators and appropriate phantoms. With today's high resolution and high-speed gamma cameras, the conditions under which a camera is used would appear to play a larger role in determining its performance than would the small differences in inherent capabilities between the cameras of various manufacturers.

摘要

相似文献

1
Methods for comparing the performance of different gamma cameras.
Semin Nucl Med. 1977 Oct;7(4):299-304. doi: 10.1016/s0001-2998(77)80038-x.
2
Performance of a PSPMT based detector for scintimammography.基于多阳极微通道板光电倍增管的乳腺闪烁成像探测器性能研究
Phys Med Biol. 2000 Mar;45(3):781-800. doi: 10.1088/0031-9155/45/3/315.
3
The variation of intrinsic spatial resolution across the UFOV of scintillation cameras.闪烁相机固有空间分辨率在有用视野范围内的变化。
Comput Med Imaging Graph. 2006 Dec;30(8):417-26. doi: 10.1016/j.compmedimag.2006.06.003. Epub 2006 Aug 8.
4
Deadtime measurements in scintillation cameras under scatter conditions simulating quantitative nuclear cardiography.在模拟定量核素心肌显像的散射条件下闪烁相机的死时间测量
J Nucl Med. 1978 May;19(5):538-44.
5
Image artifacts at high photon fluence rates in single-crystal NaI(T1) scintillation cameras.
J Nucl Med. 1978 Apr;19(4):407-13.
6
Performance evaluations of recent wide field scintillation gamma cameras.近期宽视野闪烁γ相机的性能评估。
J Nucl Med. 1978 Aug;19(8):942-7.
7
Component resolution indices for scintillation camera systems.闪烁相机系统的组件分辨率指标。
J Nucl Med. 1975 Mar;16(3):228-30.
8
Effects of asymmetric photopeak windows on flood field uniformity and spatial resolution of scintillation cameras.不对称光峰窗对闪烁相机泛光场均匀性和空间分辨率的影响。
J Nucl Med. 1986 May;27(5):706-13.
9
Hot and cold contrasts in high-resolution Tc-99m planar scintigraphy: a survey of fifty-two camera heads using the PICKER thyroid phantom.高分辨率 Tc-99m 平面闪烁照相术的冷热对比:使用 PICKER 甲状腺模型对 52 个头的摄像机进行的调查。
Phys Med. 2010;26(3):166-72. doi: 10.1016/j.ejmp.2009.06.003. Epub 2009 Jul 15.
10
Simultaneous fluoroscopic and nuclear imaging: impact of collimator choice on nuclear image quality.同步荧光透视与核成像:准直器选择对核图像质量的影响。
Med Phys. 2017 Jan;44(1):249-261. doi: 10.1002/mp.12010. Epub 2017 Jan 3.