用于氧饱和度活体视频显微镜检查的光学三倍器的研制

Development of an Optical Triplicator for intravital video microscopy of oxygen saturation.

作者信息

Mott E, Pittman R, Grant J W

机构信息

Virginia Polytechnic Institute and State University, Engineering Science and Mechanics Department, Blacksburg 24061, USA.

出版信息

IEEE Trans Biomed Eng. 1996 Nov;43(11):1116-9. doi: 10.1109/10.541253.

Abstract

The Optical Triplicator produces three copies of a portion of a microscopic image and places them side-by-side on the face of a video image tube, so that all three images can be viewed simultaneously in each video frame. The Optical Triplicator was used in an intravital microscopy assembly to obtain simultaneous images of a microvessel at three visible wavelengths selected to enable the accurate determination of oxygen saturation in microvessels of the hamster retractor muscle. An image processing system was used to obtain light intensity and optical density from video recordings made using the triplicator. Lumenal oxygen saturation profiles were determined using the measured intensity values and a published three wavelength photometric method.

摘要

光学复制器可生成微观图像一部分的三个副本,并将它们并排放置在视频显像管表面,这样在每个视频帧中都能同时查看所有三个图像。光学复制器用于活体显微镜组件中,以在三个选定的可见波长下获取微血管的同步图像,从而能够准确测定仓鼠牵张肌微血管中的氧饱和度。使用图像处理系统从使用复制器拍摄的视频记录中获取光强度和光密度。利用测量的强度值和已发表的三波长光度法确定管腔氧饱和度分布。