McGlone O, Paisley F, Shields M D, Carson D
Royal Belfast Hospital for Sick Children, Queen's University of Belfast.
Ir Med J. 1997 Jan-Feb;90(1):36-7.
Comparison of the Ames DCA 2000, a new device for the rapid measurement of HbA1c, with the established laboratory method (Dako HbA1c system) was made in 58 children routinely attending a diabetic outpatient clinic. Results showed that DCA 2000 and laboratory estimations of HbA1c were closely correlated throughout the range of values (R = 0.95). The mean difference between values was 1.83% with a 95% range of agreement of 0.23-3.4%. The DCA 2000 can be used to provide a rapid estimate of HbA1c upon which decisions on changes in treatment can be based at diabetic clinics.
在一家糖尿病门诊诊所的58名常规就诊儿童中,对用于快速测量糖化血红蛋白(HbA1c)的新设备Ames DCA 2000与既定的实验室方法(达科HbA1c系统)进行了比较。结果显示,在整个数值范围内,DCA 2000测量的HbA1c与实验室测量结果密切相关(R = 0.95)。两者测量值的平均差异为1.83%,95%的一致性范围为0.23 - 3.4%。DCA 2000可用于快速估算HbA1c,糖尿病诊所可据此做出治疗调整决策。