Calicheamicin derivatives conjugated to monoclonal antibodies: determination of loading values and distributions by infrared and UV matrix-assisted laser desorption/ionization mass spectrometry and electrospray ionization mass spectrometry.

作者信息

Siegel M M, Tabei K, Kunz A, Hollander I J, Hamann R R, Bell D H, Berkenkamp S, Hillenkamp F

机构信息

Wyeth-Ayerst Research, Lederle Laboratories, Peart River, New York 10965, USA.

出版信息

Anal Chem. 1997 Jul 15;69(14):2716-26. doi: 10.1021/ac970035q.

Abstract

Calicheamicin derivatives (MW approximately 1500) and monoclonal antibodies (MoAbs) conjugated to calicheamicin derivatives (MW approximately 150,000) were analyzed by UV-MALDI/MS, IR-MALDI/MS, and ESI/MS. These materials are potent anticancer agents. Calicheamicin derivatives and conjugates rapidly degrade upon UV irradiation but are relatively stable during IR irradiation and under ESI conditions. A unique feature of IR-MALDI/MS is a 2 times enhancement in resolution relative to UV-MALDI/MS for masses above approximately 50,000 Da resulting in a molecular ion envelope containing a series of partially resolved peaks of the calicheamicin-MoAb conjugates. The mass shift difference between the peak maxima corresponded to the mass change due to the covalent addition of calicheamicin derivatives to the monoclonal antibody. The distribution of the calicheamicin derivatives in the monoclonal antibodies was computed by deconvoluting the partially resolved peak envelope. A unique feature of the ESI mass spectra, under unit resolution conditions, is that the distribution of the carbohydrates can be well resolved for pure MoAbs and can be only partially resolved for conjugated MoAbs. Average loading values for calicheamicia derivatives when conjugated to MoAbs were computed from UV-MALDI/MS, IR-MALDI/MS, and ESI/MS data and the results compared with the average loading values obtained by UV absorption spectrometry. Very low average loading values were computed from UV-MALDI/MS data due to the degradation of the conjugated calicheamicin derivatives during the UV irradiation process. The IR-MALDI/MS average loading values, obtained with glycerol as the matrix, were consistent with the UV absorption spectrometry values for conjugates having hydrolytically stable linkers, but not when the linker contained a hydrolytically labile hydrazone. ESI/MS average loading values were generally lower than the corresponding values obtained by IR-MALDI/MS. The average loading values and distributions obtained using IR-MALDI/MS were more reliable than the corresponding ESI/MS values because the partially resolved, singly and doubly charged peaks in the IR-MALDI spectra can be mathematically deconvoluted, while the overlapping, highly multiply charged peaks of the electrospray spectra can only be partially deconvoluted.

摘要

用紫外基质辅助激光解吸电离质谱(UV-MALDI/MS)、红外基质辅助激光解吸电离质谱(IR-MALDI/MS)和电喷雾质谱(ESI/MS)对刺孢霉素衍生物(分子量约1500)以及与刺孢霉素衍生物偶联的单克隆抗体(MoAbs,分子量约150,000)进行了分析。这些物质都是强效抗癌剂。刺孢霉素衍生物及其偶联物在紫外照射下会迅速降解,但在红外照射和电喷雾条件下相对稳定。IR-MALDI/MS的一个独特之处在于,对于分子量高于约50,000 Da的物质,其分辨率相对于UV-MALDI/MS提高了2倍,从而得到一个分子离子包络,其中包含一系列刺孢霉素-MoAb偶联物的部分分辨峰。峰最大值之间的质量位移差异对应于刺孢霉素衍生物与单克隆抗体共价结合导致的质量变化。通过对部分分辨的峰包络进行去卷积计算,得出了刺孢霉素衍生物在单克隆抗体中的分布情况。在单位分辨率条件下,ESI质谱的一个独特之处在于,对于纯MoAbs,碳水化合物的分布能够很好地分辨出来,而对于偶联的MoAbs,只能部分分辨。根据UV-MALDI/MS、IR-MALDI/MS和ESI/MS数据计算了刺孢霉素衍生物与MoAbs偶联时的平均负载值,并将结果与通过紫外吸收光谱法获得的平均负载值进行了比较。由于偶联的刺孢霉素衍生物在紫外照射过程中会降解,因此从UV-MALDI/MS数据计算得出的平均负载值非常低。以甘油为基质获得的IR-MALDI/MS平均负载值,对于具有水解稳定连接子的偶联物,与紫外吸收光谱法的值一致,但当连接子含有水解不稳定的腙时则不一致。ESI/MS平均负载值通常低于通过IR-MALDI/MS获得的相应值。使用IR-MALDI/MS获得的平均负载值和分布比相应的ESI/MS值更可靠,因为IR-MALDI光谱中部分分辨的单电荷和双电荷峰可以通过数学方法进行去卷积,而电喷雾光谱中重叠的高度多电荷峰只能部分去卷积。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验