Suppr超能文献

Nonuniformity intrinsic flood artifact caused by point-source syringe needle.

作者信息

Hackett M T, Magoun S L, Thompson R B

机构信息

Nuclear Medicine Service, Department of Veterans Affairs Medical Centers, Lexington, Kentucky 40511-1093, USA.

出版信息

J Nucl Med Technol. 1997 Mar;25(1):41-3.

PMID:9239602
Abstract

UNLABELLED

The purpose of this study was to investigate the cause of a nonuniformity artifact seen on an intrinsic flood.

METHODS

Intrinsic floods were acquired on three different cameras using a point source in a 1-cc syringe, with and without a needle attached. Floods were obtained with the syringe placed perpendicular and parallel to the camera face.

RESULTS

A nonuniformity artifact was evident if a syringe with a needle was placed perpendicular to the crystal face.

CONCLUSION

If a syringe is used as a point source during the acquisition of an intrinsic quality control flood, the needle should not be placed perpendicular to the camera face.