Hackett M T, Magoun S L, Thompson R B
Nuclear Medicine Service, Department of Veterans Affairs Medical Centers, Lexington, Kentucky 40511-1093, USA.
J Nucl Med Technol. 1997 Mar;25(1):41-3.
The purpose of this study was to investigate the cause of a nonuniformity artifact seen on an intrinsic flood.
Intrinsic floods were acquired on three different cameras using a point source in a 1-cc syringe, with and without a needle attached. Floods were obtained with the syringe placed perpendicular and parallel to the camera face.
A nonuniformity artifact was evident if a syringe with a needle was placed perpendicular to the crystal face.
If a syringe is used as a point source during the acquisition of an intrinsic quality control flood, the needle should not be placed perpendicular to the camera face.