Shanthi S, Singh M
Biomedical Engineering Division, Indian Institute of Technology, Madras, India.
Med Biol Eng Comput. 1997 May;35(3):253-8. doi: 10.1007/BF02530046.
Based on the principle of backscattering of laser radiation from tissues, a non-invasive PC-AT based reflectance imaging technique is developed. The laser beam from a semiconductor laser operating at 670 nm is guided to the tissue site by an optical fibre. The backscattered radiation is collected by another fibre placed in the same probe, and is detected by a photodiode-amplifier assembly. This probe is moved manually over the organs under observation, and the data after the ADC, interpolation and median filtering are displayed in the form of reflectance image of the organ along with grey scale. By this technique images of the human hands and forearms are obtained, which depend on the variations in their colour, composition and blood flow. A comparison is made with perfusion images, obtained by a Periflux laser Doppler flowmeter. These show that the reflectance images provide greater details of the tissue structure than the perfusion images.