与心率相关的束支传导阻滞期间植入式心脏复律除颤器感知过度

Implantable cardioverter defibrillator oversensing during periods of rate-related bundle branch block.

作者信息

Coyne R F, Man D C, Sarter B H, Schwartzman D, Callans D J, Marchlinski F E, Gottlieb C D

机构信息

Allegheny University of the Health Sciences, Philadelphia, Pennsylvania 19129, USA.

出版信息

J Cardiovasc Electrophysiol. 1997 Jul;8(7):807-11. doi: 10.1111/j.1540-8167.1997.tb00839.x.

Abstract

A patient with an implantable cardioverter defibrillator (ICD) and a dual chamber pacemaker experienced inappropriate ICD therapies only during periods of rate-dependent right bundle branch block. Analysis of both stored and real-time ICD electrograms was critical to correctly diagnosing the problem and offering a solution.

摘要

一名植入式心脏复律除颤器(ICD)和双腔起搏器的患者仅在心率依赖性右束支传导阻滞期间出现不适当的ICD治疗。对存储的和实时的ICD心电图进行分析对于正确诊断问题并提供解决方案至关重要。