Artiss J D, Yang W C, Harake B, Capellari E, Kretch C, Eisenbrey A B, Zak B
Department of Pathology, Wayne State University School of Medicine, Detroit, Michigan 48201, USA.
Am J Clin Pathol. 1997 Sep;108(3):269-74. doi: 10.1093/ajcp/108.3.269.
We describe a modification of a previously described serum iron procedure applied to the Bayer DAX48 (Bayer Diagnostics, Tarrytown, NY) automated chemistry analyzer. The iron-ligand used in this assay, 2-(5-nitro-2-pyridylazo)-5-(N-propyl-N-sulfopropylamine) phenol (nitro-PAPS), has a molar absorptivity of 94,000 L mol(-1) cm(-1), which is three to four times more sensitive than the more commonly used ligands. The increased sensitivity of the iron-ligand complex facilitates modification of a Ferene S method that requires a smaller sample volume while it maintains the precision of the assay. Because the reagent does not contain ascorbate, the "onboard" stability has been increased to more than 4 weeks. The reagent seems to be quite insensitive to icterus and hemolysis. Furthermore, the interference of turbidity caused by triglycerides, abnormal proteins, or fibrinogen, present in samples from patients undergoing anticoagulant therapy, seems to have been eliminated.
我们描述了一种对先前所述血清铁检测方法的改进,该方法应用于拜耳DAX48(纽约塔里敦拜耳诊断公司)自动化学分析仪。本检测中使用的铁配体2-(5-硝基-2-吡啶偶氮)-5-(N-丙基-N-磺丙胺)苯酚(硝基-PAPS)的摩尔吸光系数为94,000 L·mol⁻¹·cm⁻¹,比更常用的配体灵敏三到四倍。铁配体络合物灵敏度的提高有助于对Ferene S方法进行改进,该方法所需样本量更小,同时保持检测的精密度。由于该试剂不含抗坏血酸,“机载”稳定性已提高到4周以上。该试剂似乎对黄疸和溶血相当不敏感。此外,抗凝治疗患者样本中甘油三酯、异常蛋白质或纤维蛋白原引起的浊度干扰似乎已被消除。