Suppr超能文献

Intrinsic Dacron graft failure 19 years post-implantation.

作者信息

Nunn D B, Pourdeyhimi B

机构信息

University of Florida, Health Science Center, Jacksonville, USA.

出版信息

Cardiovasc Surg. 1997 Jun;5(3):333-8. doi: 10.1016/s0967-2109(97)00028-8.

Abstract

Intrinsic failure of a knitted Dacron aortic bifurcation graft 19 years after implantation is reported. Failure was due to inability of the dilated and deteriorated left graft limb to hold anastomotic sutures. Numerous transverse cracks and breaks were detected in the Dacron filaments, in addition to small holes throughout the fabric structure, and a severe reduction in the yarn's tensile strength. Criteria are defined for establishing a correct diagnosis, and causative factors, incidence, and clinical management are discussed.