Field J R, Hearn T C, Caldwell C B
Equine Research Centre, Guelph, Ontario, Canada.
J Orthop Trauma. 1997 Jul;11(5):368-73. doi: 10.1097/00005131-199707000-00013.
To determine the effects of bone morphology on the interfacial mechanics of bone plate fixation.
A comparison of two bone plate designs using measures of contact area and force in three conditions of bone type.
The influence of bone plate design, bone type, and implant location on interfacial characteristics of contact area and average force were evaluated using pressure-sensitive film in cadaveric bone. Data were quantitated using computer-assisted image analysis.
Means of the two dependent variables in each of the three conditions were compared using analysis of variance.
In human femora and equine third metacarpi, there were no apparent differences in interface contact area attributed to bone plate design, although there were significant differences for interface contact force. In humeri, significant differences in both interface contact area and force were found between bone plate designs.
Results show that bone surface topography and implant location are major determinants of the interfacial elements of contact area and average force between a plate and the bone to which it is applied.
确定骨形态对接骨板固定界面力学的影响。
在三种骨类型条件下,通过比较两种接骨板设计的接触面积和力的测量值进行研究。
使用压敏膜在尸体骨中评估接骨板设计、骨类型和植入位置对接触面积和平均力等界面特征的影响。数据通过计算机辅助图像分析进行定量。
使用方差分析比较三种条件下每个因变量的均值。
在人股骨和马第三掌骨中,尽管界面接触力存在显著差异,但接骨板设计对界面接触面积无明显影响。在肱骨中,接骨板设计在界面接触面积和力方面均存在显著差异。
结果表明,骨表面形态和植入位置是接骨板与其所应用骨之间接触面积和平均力等界面要素的主要决定因素。