Mitchell S J, Willcox T, Gorman D F
Diving and Hyperbaric Medicine Unit, Royal New Zealand Navy Hospital, Devonport, Auckland, New Zealand.
Perfusion. 1997 Sep;12(5):325-33. doi: 10.1177/026765919701200508.
We have previously shown significant bubble formation in Medtronic Maxima hard-shell venous reservoirs (HSVRs). In the present study, we not only investigated the mechanism of this bubble formation, but also the extent of bubble clearance by membrane oxygenators and arterial line filters. In addition, we also compared the performance of five HSVRs with respect to bubble formation and venous air filtration. Salvaged clinical CPB circuits containing different HSVRs were studied by downstream Doppler monitoring under fixed flow-decreasing volume, fixed volume-increasing flow, and entrained venous air conditions. Bubbles formed in the Medtronic Maxima top entry HSVR at volumes below 800 ml and flows above 3.5 l min-1, and were incompletely removed by a membrane oxygenator and arterial line filter. Decreased bubbling was seen when the reservoir atmosphere was flushed with CO2, suggesting that these bubbles formed in a fountain at the venous inflow. The Medtronic Maxima Forte HSVR formed significantly fewer bubbles at low volumes, and filtered venous air effectively. Negligible bubble formation occurred in the Sorin, Terumo, or Baxter reservoirs. The minimum recommended operating volume for the Medtronic Maxima top entry reservoir should be reset at 600 ml and this device should always be used with an arterial filter. Bubble formation is substantially reduced in the new Medtronic Maxima Forte HSVR and this device is a good filter for venous air.
我们之前已证明美敦力Maxima硬壳静脉储血器(HSVRs)中会形成大量气泡。在本研究中,我们不仅研究了这种气泡形成的机制,还研究了膜式氧合器和动脉管路过滤器清除气泡的程度。此外,我们还比较了五种HSVRs在气泡形成和静脉空气过滤方面的性能。通过在固定流量减少体积、固定体积增加流量和夹带静脉空气条件下的下游多普勒监测,研究了包含不同HSVRs的回收临床体外循环回路。在美敦力Maxima顶部入口HSVR中,当体积低于800 ml且流量高于3.5 l min-1时会形成气泡,并且膜式氧合器和动脉管路过滤器不能完全清除这些气泡。当储血器内气体用二氧化碳冲洗时,气泡形成减少,这表明这些气泡是在静脉入口处的喷泉中形成的。美敦力Maxima Forte HSVR在低体积时形成的气泡明显较少,并且能有效过滤静脉空气。在索林、泰尔茂或百特储血器中几乎没有气泡形成。美敦力Maxima顶部入口储血器的最低推荐操作体积应重新设定为600 ml,并且该设备应始终与动脉过滤器一起使用。新型美敦力Maxima Forte HSVR中的气泡形成显著减少,并且该设备是一种良好的静脉空气过滤器。