Evaluation of bedside prothrombin time and activated partial thromboplastin time measurement by coagulation analyzer CoaguCheck Plus in various clinical settings.

作者信息

Ruzicka K, Kapiotis S, Quehenberger P, Handler S, Hornykewycz S, Michitsch A, Huber K, Clemens D, Susan M, Pabinger I, Eichinger S, Jilma B, Speiser W

机构信息

Clinical Institute of Medical and Chemical Laboratory Diagnostics, University of Vienna, Austria.

出版信息

Thromb Res. 1997 Sep 1;87(5):431-40. doi: 10.1016/s0049-3848(97)00159-x.

Abstract

In the present study CoaguCheck Plus (CCP), a coagulation test system using whole blood, was evaluated with respect to its comparability with widely distributed conventional routine coagulation assays. A correlation of r = 0.997 (p < 0.0001) was found between INR of CCP-prothrombin time (CCP-PT) and Thrombotest (KC-1 analyzer) in patients on oral anticoagulant therapy. A correlation of r = 0.899 (p < 0.001) between CCP-aPTT and Actin ES aPTT (STA analyzer) was found in heparinized patients. Impaired hepatic hepatic coagulation factor synthesis in liver cirrhosis patients was detected by CCP-PT with a sensitivity of 0.75 and by Normotest (STA analyzer) with a sensitivity of 0.92. Those patients with normal CCP-PT values and liver disease had, only mild reductions (> 30% of normals) in coagulation factors II, V, VII or X. CCP-aPTT was also performed in patients with a deficiency in the so called endogenous coagulation factors VIII, IX, XI and XII. CCP-aPTT showed a sensitivity similar to that of Actin FS aPTT in the detection even of mild deficiencies in factors VIII, IX and XII; factor XI deficiency was however detected only in patients with severe (< 12% of normals) disease; lupus anticoagulants were detected with a high sensitivity.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验