Suppr超能文献

带有磁悬浮叶轮的离心泵连续运行超过1年。

More than 1 year continuous operation of a centrifugal pump with a magnetically suspended impeller.

作者信息

Nojiri C, Kijima T, Maekawa J, Horiuchi K, Kido T, Sugiyama T, Mori T, Sugiura N, Asada T, Shimane H, Nishimura K, Ban T, Akamatsu T, Ozaki T, Ito H, Suzuki M, Akutsu T

机构信息

Terumo Corp., R&D Center, Kanagawa, Japan.

出版信息

ASAIO J. 1997 Sep-Oct;43(5):M548-52.

PMID:9360103
Abstract

The authors have been developing a centrifugal pump with a magnetically suspended impeller (MSCP) designed for total artificial heart and long-term ventricular support. The MSCP consists of a magnetic bearing, an impeller and housing, and a driving motor. The impeller is suspended by a magnetic bearing, therefore providing contact free rotation of the impeller inside the pump. This study was designed to evaluate long-term durability and nonthrombogenicity of the MSCP in a chronic sheep model. The blood contacting surfaces of the pump and conduits were completely modified by a heparin immobilization technique (Hepaface). The MSCP was placed paracorporeally as a left heart bypass between left ventricle and descending aorta in three adult sheep. Coumadin was given orally to maintain prothrombin time at 15-20 sec. The coagulation and hematologic parameters, including plasma free hemoglobin, were periodically monitored throughout the experiment. Under daily movement in the cage, the pump could produce average flow rates of 3-6 L/min (50-100 ml/kg) at 1,700-2,000 rpm. Although the arterial pulse contour decreased, there was no physiologic deterioration. The axial impeller excursion monitored by a position sensor was < 25 microns. Plasma free hemoglobin level remained at < 5 mg/dl throughout the experiment. There was no increase in the motor current, which indicates no massive thrombus formation around the impeller. One experiment was terminated at 70 days due to Hall sensor dysfunction of the motor. The retrieved pump was entirely free from thrombus formation. There was no detectable thrombus formation inside the pump or the inflow and outflow conduits. Hematologic, renal, and hepatic parameters remained within the normal range throughout the experiment. The other two sheep have survived for more than 395 and 41 days without major complication. These studies demonstrated that the MSCP has significant potential for long-term use.

摘要

作者一直在研发一种带有磁悬浮叶轮的离心泵(MSCP),该泵专为全人工心脏和长期心室支持而设计。MSCP由磁轴承、叶轮和外壳以及驱动电机组成。叶轮由磁轴承悬浮,因此可在泵内实现叶轮的无接触旋转。本研究旨在评估MSCP在慢性绵羊模型中的长期耐久性和非血栓形成性。泵和导管的血液接触表面通过肝素固定技术(Hepaface)进行了完全改性。在三只成年绵羊中,将MSCP作为左心旁路体外置于左心室和降主动脉之间。口服香豆素以将凝血酶原时间维持在15 - 20秒。在整个实验过程中定期监测凝血和血液学参数,包括血浆游离血红蛋白。在笼中日常活动的情况下,该泵在1700 - 2000转/分钟时可产生3 - 6升/分钟(50 - 100毫升/千克)的平均流速。尽管动脉脉搏轮廓降低,但未出现生理恶化。由位置传感器监测的轴向叶轮偏移<25微米。在整个实验过程中,血浆游离血红蛋白水平保持在<5毫克/分升。电机电流没有增加,这表明叶轮周围没有大量血栓形成。由于电机的霍尔传感器功能障碍,一项实验在70天时终止。回收的泵完全没有血栓形成。在泵内以及流入和流出导管内未检测到血栓形成。在整个实验过程中,血液学、肾脏和肝脏参数均保持在正常范围内。另外两只绵羊存活了超过395天和41天,没有出现重大并发症。这些研究表明,MSCP具有长期使用的巨大潜力。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验